Face Mask and Respirator CE
standards/labelling
Respirators
(FFP2, FFP3) and medical face masks (Type I, II & IIR) correspond to 2
different EU standards and regulations:
1.Medical
Face Masks Type I, II & IIR fall under the Medical Device Directive/Regulation
(MDD/MDR)
A
medical face mask is used to protect the patient from micro-organisms in the
wearer's (healthcare professional's) breath.
Surgical
face masks of Type IIR also protect the wearer against liquid splashes>
120mmHg (average systolic arterial blood pressure i.e.: possible strength of a
blood splash in case of rupture in small arteries).
Medical
face masks should be marked with a CE-mark showing that it conforms with the
requirements of the Medical Device Directive 93/42 (MDR as from end May 2020)
as Class 1 medical device.
The
CE mark must only be displayed on the packaging (it is not accompanied by the
registration number of a Notified Body).
EN
14683 is the European standard setting the construction and performance
requirements, and test methods for medical and surgical masks.
Main
performance requirements according to standard:
• Bacterial Filtration Efficiency
Determines
the filtration efficiency of the face mask: the higher the BFE %, the better
the protection.
Classification:
• TYPE I
= ≥ 95 %
• TYPE II = ≥ 98 %
• Breathing Resistance
Determines
the airflow resistance of the surgical mask: the lower the differential
pressure, the higher the breathing comfort.
Classification:
• TYPE I & II (non splash resistant)
= < 29.4 Pa/cm2
• TYPE IR & IIR (splash resistant) =
< 49.0 Pa/cm2
• Splash Resistance
Determines
the resistance of the mask to potentially contaminated fluid splashes.
Classification:
• TYPE I & TYPE II: not applicable
• TYPE IR & TYPE IIR: >120 mmHg*
2.FFP
respirators are falling under the Personal Protective Equipment (PPE)
Regulation:
FFP3
stands for:
·
FF
= filtering facepiece (filtering half mask)
·
P
= particles
·
3
= class (where 3 is the class that has the highest protection effect)
A
respirator is intended to protect the wearer (healthcare personnel) against
airborne particles (e.g.: fine dust particles, but also possibly virus
particles).
All
personal protective equipment, such as respiratory protection, must be clearly
marked on the product according to the regulations in force.
In
the new European regulation 2016/425 on Personal Protective Equipment,
respiratory protection “Category III covers exclusively risks that can lead to
very serious consequences, such as death or permanent health damage, e.g. with
regard to hazardous substances and mixtures ”, which is the strictest category.
EN
149 + A1: 2009 is a European standard for respiratory protection, filtering
half masks for protection against particles. The standard sets requirements for
both the product and the manufacturer.
Requirements
according to standard
·
Labeling
of product according to standard
Tests
to be performed according to the standard are:
·
Total
leakage
·
Filter
penetration
·
Leakage
around the face
·
Leakage
in the valve
·
Flame
safety
·
Respiratory
resistance
·
Exposure
test = extended loading; a lab test to ensure that the filter material does not
start to leak
·
Clogging
test (= clogging test); a voluntary test of disposable breathing protection.
Marked with "D" (D = Dolomite dust, which is the substance used in
the test)
Respiratory
protection is divided into three different classes; FFP1, FFP2, FFP3 according
to:
·
Efficiency
(E%)
·
Total
leakage (P%)
·
Protection
factor
The
respirators should be marked with;
·
Manufacturer
name and item number (e.g.: 3M 1883+)
·
Which
standard and class it meets (e.g.: EN 149 + A1: 2009, FFP3: 2001)
·
CE-Mark
followed by 4 digits (e.g.: CE0086) showing that it is approved by a notified
body as conform with the requirements of the strictest PPE category III.
The
CE marking must be clearly marked on each individual product and the
instruction manual must be in local language (local market).
Some products may be dual
labelled both as Medical Face Mask (Type IIR) AND FFP3 Respirator. But they are
quite rare (most respirators are sold as dust masks PPE).
In any case, it is still
worth to consider that product even if single labelled (either as medical face
mask or as protective respirator) will still provide some level of protection
for both the user and the patient (or people around them).
|